7102
User's Guide
9-4
9.3 Scan
The scan rate can be set and enabled so that when the set-point is changed the
bath heats or cools at a specified rate (degrees per minute) until it reaches the
new set-point. With the scan disabled the bath heats or cools at the maximum
possible rate.
9.3.1 Scan Control
The scan is controlled with the scan on/off function that appears in the main
menu after the set-point function.
Sc=OFF Scan function off
Press “UP” or “DOWN” to toggle the scan on or off.
Sc=On Scan function on
Press “SET” to accept the present setting and continue.
Accept scan setting
9.3.2 Scan Rate
The next function in the main menu is the scan rate. The scan rate can be set
from .1 to 99.9 °C/min. The maximum scan rate however is actually limited by the
natural heating or cooling rate of the instrument. This is often less than
100 °C/min, especially when cooling.
The scan rate function appears in the main menu after the scan control function.
The scan rate units are in degrees C per minute.
Sr= 10.0 Scan rate in °C/min
Press “UP” or “DOWN” to change the scan rate.
Sr= 2.0 New scan rate
Press “SET” to accept the new scan rate and continue.
Accept scan rate
9.4 Temperature Display Hold
The 7102 has a display hold function which allows action of an external switch to
freeze the displayed temperature and stop the set-point from scanning. This is
useful for testing thermal switches and cut-outs. The instrument must be
powered off before attaching thermal switches or cut-outs. This section explains
the functions available for operating the temperature hold feature. An example
follows showing how to set up and use the hold feature to test a switch.