3-1
Chapter 3
Specifications and Environmental
Conditions
3.1 Specifications
Table 2. Specifications
Range 5 to 125 °C (23 to 257 °F)
Accuracy ±0.25 °C
Stability ±0.015 °C at 5 °C (oil, 5010)
±0.03 °C at 121 °C (oil, 5010)
Uniformity ±0.02 °C
Resolution 0.01 °C/F
Operating Temperature 5 to 45 °C (41 to 113 °F)
Heating Time 25 °C to 100 °C (77 °F to 212 °F): 30 minutes
Cooling Time 25 °C to 0 °C (77 °F to 32 °F): 30 minutes
Well Size 2.5" dia. x 5.5" deep (64 x 139 mm)
(access opening is 1.9" [48 mm] in diameter )
Exterior Dimension 12" H x 7.2" W x 9.5" D (31 cm x 18 cm x 24 cm)
Weight 15 lb. (6.8 kg) with fluid
Power 115 VAC ( ±10 %), 1.8 A, or 230 VAC ( ±10 %), 0.9
A, switchable, 50/60 Hz, 200 W
Readout Switchable °C or °F
Controller Digital controller with data retention
Thermal Electric Devices (TED) 150 W
Cooling Fan and Thermal Electric Devices (TED)
Fault Protection Sensor burnout and short protection
Safety Overvoltage (Installation) Category II, Pollution
Degree 2 per IEC 61010-01
Fuse Rating 115 V: 250 V 3A SB (slow blow)
230 V: 250 V 1.6 A T (time delay)