Controller Operation
9.8 Controller Configuration 9
9-9
9.8.1.1 High Limit
The High Limit Parameter adjusts the upper set-point temperature. The factory
default and maximum temperature are set to 126 °C. For safety, a user can
adjust the HL down so the maximum temperature set-point is restricted.
HL High Limit parameter
Press “SET” to enable adjustment of HL
HL Flashes “HL” and then displays the setting
H=126 Current HL setting
Adjust the HL parameter using “UP” or “DOWN”
H=90 New HL setting
Press “SET “to accept the new temperature limit.
9.8.1.2 Stir Speed
The Stir Speed parameter adjusts stirrer motor speed. The factory default is 20.
Str SP Flashes “Str Sp” and then displays the setting
0 Current Stir Speed setting
To change the stir speed press “UP” or “DOWN”.
16 New Stir Speed setting
Press “SET” to accept the new Stir Speed.
The stir motor speed needs to be varied for best stability. Table 3 on page 25
shows nominal settings for several fluids.
9.8.2 Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
SErIAL Serial RS-232 interface parameters menu
The serial interface parameters menu contains parameters which determine the
operation of the serial interface. These controls only apply to instruments fitted
with the serial interface. The parameters in the menu are — baud rate, sample
period, duplex mode, and linefeed. Press “UP” to enter the menu.