FLUKE 7102-156 Liquid Bath Calibrator, -5 to 125°C

User Manual for the Fluke 7102-156 Liquid Bath Calibrator, -5 to 125°C - Page 13

For 7102-156.

PDF File Manual, 68 pages, Read Online | Download pdf file

7102-156 photo
Loading ...
Loading ...
Loading ...
Before You Start
1.2 Safety Information1
1-3
Before initial use, or after transport, or after storage in
humid or semi-humid environments, or anytime the
instrument has not been energized for more than 10 days,
the instrument needs to be energized for a “dry-out” period
of 2 hours before it can be assumed to meet all of the safety
requirements of the IEC 61010-1. If the product is wet or has
been in a wet environment, take necessary measures to
remove moisture prior to applying power such as storage in
a low humidity temperature chamber operating at 50 °C for 4
hours or more.
The instrument is intended for indoor use only.
Lift the instrument by the handle provided to move the
instrument. DO NOT move the instrument until the display
reads less than 25 °C (77 °F) and the unit has been drained
or the Transport Lid installed.
BURN HAZARD
ALWAYS ensure the instrument is COOL before closing the
instrument for storage.
DO NOT touch the well access surface of the unit.
DO NOT mix water and oil when exceeding temperatures of
90 °C (194 °F).
DO NOT mix water and oil when exceeding temperatures of
90 °C (194 °F). The temperature of the well access is the
same as the actual temperature shown on the display, e.g. if
the unit is set at 125 °C and the display reads 125 °C, the
well is at 125 °C.
Ensure the power cord is positioned in such a way as it
cannot contact hot surfaces or temperature probes. Always
inspect power cord before use for any damage to the
insulation due to contact with hot surfaces, cuts or
abrasions.
The top sheet metal of the instrument may exhibit extreme
temperatures for areas close to the well access.
DO NOT turn off the unit at temperatures higher than 100 °C.
This could create a hazardous situation. Select a set-point
less than 100 °C and allow the unit to cool before turning it
off.
DO NOT remove the fluid at high temperatures. The fluid will
be the same temperature as the display temperature.
DO NOT operate near flammable materials. Extreme
temperatures could ignite the flammable material.
Use of this instrument at HIGH TEMPERATURES for
extended periods of time requires caution.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com