Controller Operation
9.2 Temperature Set-point 9
9-3
100.00C Well temperature in degrees Celsius
Access set-point memory
1. 25 Set-point memory 1, 25 °C currently used
To change the set-point memory press “UP” or “DOWN”.
4. 125. New set-point memory 4, 125 °C
Press “SET” to accept the new selection and access the set-point value.
Accept selected set-point memory
9.2.2 Set-point Value
The set-point value may be adjusted after selecting the set-point memory and
pressing “SET”.
4 125. Set-point 4 value in °C
If the set-point value is correct, hold “EXIT” to resume displaying the well
temperature. Press “UP” or “DOWN” to adjust the set-point value.
125.00 New set-point value
When the desired set-point value is reached press “SET” to accept the new value
and access the temperature scale units selection. If “EXIT” is pressed instead of
“SET”, any changes made to the set-point are ignored.
Accept new set-point value
9.2.3 Temperature Scale Units
The temperature scale units of the controller can be set by the user to degrees
Celsius ( °C) or Fahrenheit ( °F). The selected units are used in displaying the
well temperature, set-point, and proportional band.
Press “SET” after adjusting the set-point value to change display units.
Un= C Scale units currently selected
Press “UP” or “DOWN” to change the units.
Un= F New units selected