11-1
Chapter 11
Test Probe Calibration
Note:
This procedure is to be considered a general guideline. Each
laboratory should write their own procedure based on their
equipment and their quality program. Each procedure should be
accompanied by an uncertainty analysis also based on the
laboratory’s equipment and environment.
For optimum accuracy and stability, allow the calibrator to warm up for 25
minutes after power-up and then allow adequate stabilization time after reaching
the set-point temperature. After completing operation of the calibrator, allow the
well to cool by setting the temperature to 25 °C for one-half hour before switching
the power off.
11.1 Calibrating a Single Probe
Insert the probe to be calibrated into the well of the bath. Best results are
obtained with the probe inserted to the full depth of the well. Once the probe is
inserted into the well, allow adequate stabilization time to allow the test probe
temperature to settle as described above. Once the probe has settled to the
temperature of the well, it may be compared to the calibrator display
temperature. The display temperature should be stable to within 0.1 °C degree
for best results.