7102
User's Guide
9-12
9.8.3.1 R0
This probe parameter refers to the resistance of the control probe at 0 °C. The
value of this parameter is set at the factory for best instrument accuracy.
9.8.3.2 ALPHA
This probe parameter refers to the average sensitivity of the probe between 0
and 100 °C. The value of this parameter is set at the factory for best instrument
accuracy.
9.8.3.3 DELTA
This probe parameter characterizes the curvature of the resistance-temperature
relationship of the sensor. The value of this parameter is set at the factory for
best instrument accuracy.
9.8.3.4 C0 and CG
These parameters calibrate the accuracy of the bath set-point. These are
programmed at the factory when the bath is calibrated. DO NOT alter the value
of these parameters. If the user desires to calibrate the bath for improved
accuracy then calibrate R0 and ALPHA according to the procedure in Section 12.
9.8.3.5 rCAL
DO NOT adjust this parameter. It is for factory use only.