Caution:
Never introduce any foreign material into the well.
11.2 Stabilization and Accuracy
The stabilization time of the Micro-Bath depends on the conditions and
temperatures involved. Typically the test well will be stable to 0.1 °C within 10
minutes of reaching the set-point temperature. Ultimate stability will be achieved
30 minutes after reaching the set temperature.
Inserting a cold probe into a well requires another period of stabilizing depending
on the magnitude of the disturbance and the required accuracy. For example,
inserting a.25 inch diameter room temperature probe at 200 °C takes 5 minutes
to be within 0.1 °C of its settled point and takes 10 minutes to achieve maximum
stability.
Speeding up the calibration process can be accomplished by knowing how soon
to make the measurement. Typical measurements should be made at the desired
temperatures with the desired test probes to establish these times.