8-1
Chapter 8
General Operation
8.1 Changing Display Units
The 7102 can display temperature in Celsius or Fahrenheit.
The temperature units are shipped from the factory set to Celsius.
To change to Fahrenheit or back to Celsius there are two ways:
1. Press “SET” and “UP” simultaneously. This will change the display units.
2. Press the “SET” key three times from the temperature display to show
Un = C
or
Press the “UP” or “DOWN” key to change units.
Press “SET” to store changes.
8.2 Switching to 230V Operation
The 7102 is switchable from 115 VAC to 230 VAC 50/60 Hz. Switching the
voltage can change the calibration, so the unit should be calibrated after
changing the input voltage.
To change from 115 VAC to 230 VAC:
Unplug the unit.
With a small straight slot screwdriver remove the fuse holder located on the
back of the bath. Replace the two fuses (3 amp 250 V) with 1.6 amp 250 V
fuses.
Replace the fuse holder with the “230V” in the display window.
Note:
Use 3 amp fuses for 115 V and 1.6 amp for 230V only. DO NOT
PLUG THE UNIT INTO 230 V IF THE FUSE HOLDER READS 115 V.