Chapter 16 Display Control RIGOL
MSO5000 User Guide 16-3
To Set the Waveform Intensity
Press Intensity, then rotate the multifunction knob at a small step value or use
the pop-up numeric keypad to set the waveform intensity. Note that in the non-menu
operation mode, you can also rotate the multifunction knob
to adjust the
waveform brightness of the channel. The default is 50%, and the range available is
from 1% to 100%.
To Set the Screen Grid
Press Grid, and rotate the multifunction knob to set the grid type and then press
down the knob to select it. You can also press Grid continuously or enable the touch
screen to select it. You can select "FULL", "HALF", "NONE", or "IRE".
âš« FULL: turns the background grid and coordinate on.
âš« HALF: turns the background grid off.
âš« NONE: turns the background grid and coordinate off.
âš« IRE: only available for testing the video signal (i.g. the trigger type is set to
"Video" trigger, and the vertical scale is set to 140 mV).
When "IRE" is selected, the screen displays the vertical graticule in IRE and
marks it at the left side of the screen. Its range is from -40 IRE to +100 IRE.
Meanwhile, 0.35 V and 0.7 V levels are displayed at the right side of the screen,
as shown in the following figure.
To Set the Grid Brightness
Press Brightness, then rotate the multifunction knob at a small step value or
use the pop-up numeric keypad to set the grid intensity. The default is 30%, and the