Rigol MSO5354 oscilloscope

MSO5000 user manual - Page 139

For MSO5354.

PDF File Manual, 372 pages, Read Online | Download pdf file

MSO5354 photo
Loading ...
Loading ...
Loading ...
Chapter 5 To Trigger the Oscilloscope RIGOL
MSO5000 User Guide 5-55
Figure 5-24 Trigger Output Connector
Press Utility System AUX Out to select "TrigOut”. A signal which reflects the
current capture rate of the oscilloscope can be output from the [TRIG OUT]
connector each time a trigger is generated by the oscilloscope. If this signal is
connected to a waveform display device to measure the frequency, the measurement
result is the same as the current capture rate.
Note: If you press Utility System AUX Out to select "PassFail", then in the
pass/fail test, when a pass/fail event is detected, the oscilloscope outputs a pulse
from the rear-panel [TRIG OUT] connector.
Trigger Output Connector