Source Selection
Click or tap the
Source drop-down button to select CH1-CH4. For details, refer to
Trigger Source
. The selected trigger source is indicated in the trigger information
label at the top of the screen.
Only when you select the channel that has signal inputs as the trigger source, can you
obtain a stable trigger.
Polarity
In the
Polarity item, select the desired polarity: positive polarity ( ) or negative
polarity (
).
Trigger Condition
Set the trigger condition in the
When item.
• When you select "Positive" for polarity, ">" for trigger condition, the oscilloscope
triggers when the positive pulse width of the input signal is greater than the
specified pulse width.
• When you select "Positive" for polarity, "<" for trigger condition, the oscilloscope
triggers when the positive pulse width of the input signal is smaller than the
specified pulse width.
• When you select "Positive" for polarity, "< >" for trigger condition, the
oscilloscope triggers when the positive pulse width of the input signal is greater
than the specified lower limit of pulse width and smaller than the specified
upper limit of pulse width.
• When you select "Negative" for polarity, ">" for trigger condition, the
oscilloscope triggers when the negative pulse width of the input signal is greater
than the specified pulse width.
• When you select "Negative" for polarity, "<" for trigger condition, the
oscilloscope triggers when the negative pulse width of the input signal is smaller
than the specified pulse width.
• When you select "Negative" for polarity, "< >" for trigger condition, the
oscilloscope triggers when the negative pulse width of the input signal is greater
than the specified lower limit of pulse width and smaller than the specified
upper limit of pulse width.
Pulse Width Setting
• In the When menu, when ">" or "<" is selected, click or tap the input field of
Upper or lower to set the upper limit value or the lower limit value with the
pop-up numeric keypad. You can also use the corresponding multipurpose knob
to set the value. The pulse width available is from 1 ns to 10 s.
Triggering the Oscilloscope
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO800 User Guide