non-compensated or poorly compensated probe may cause measurement errors. The
compensation procedure is as follows:
1. Perform Step 1, 2, 3 and 4 in Function Inspection.
2. Check the displayed waveforms and compare them with the waveforms shown in
Figure 4.8
.
Over compensated Under compensatedPerfectly compensated
Figure 4.8 Probe Compensation
3. Use the probe compensation adjustment tool provided in the accessories to adjust
the low-frequency compensation adjustment hole on the probe until the displayed
waveform is consistent with the "Perfectly compensated" waveform shown in the
above figure.