between two neighboring rising edges), and the idle time should be within the range
between P and M (P < Idle Time < M). Wherein, M is the time between the first rising
edge and its previous rising edge; P is the maximum time between the rising edges
that participate in counting.
P
M
Edge Type=
Edge Num=2
P<Idle Time<M
Figure 8.22 Nth Edge Trigger
Trigger Type
Click or tap the
Type drop-down button to select "Nth Edge".
Figure 8.23 Nth Edge Trigger Setting Menu
After a trigger type is selected, the current trigger setting information (including
trigger type, trigger source, and trigger level) is displayed in the trigger information
label at the top of the screen, as shown in the figure below. The information will
change based on the trigger settings.