Run Control
Press the Run/Stop or Single button on the front panel to run or stop the sampling
system of the scope.
When the Run/Stop button is green, the oscilloscope is running, that is,
acquiring data when trigger conditions are met. To stop acquiring data, press
the Run/Stop button. When stopped, the last acquired waveform is displayed.
When the Run/Stop button is red, data acquisition is stopped. A red "Stop" is
displayed next to the trademark logo in the status line at the top of the display.
To start acquiring data, press Run/Stop.
To capture and display a single acquisition (whether the oscilloscope is
running or stopped), press Single. The Single run control lets you view single-
shot events without subsequent waveform data overwriting the display. Use
Single when you want maximum memory depth for pan and zoom.
When you press Single, the display is cleared, the trigger mode is temporarily set to
Normal (to keep the oscilloscope from auto- triggering immediately), the trigger
circuitry is armed, the Single key is illuminated, and the oscilloscope waits until a
user defined trigger condition occurs before it displays a waveform.
When the oscilloscope triggers, the single acquisition is displayed and the
oscilloscope is stopped (the Run/Stop button is illuminated in red).
Press Single again to acquire another waveform.