Manual - Page 212

For SDS2202X-E.

Loading ...
Loading ...
Loading ...
6. Press the Mask Setting softkey to enter the MASK function menu.
7. Press the X Mask or Y Mask softkey; and then turn the Universal Knob to select
the desired value. The range is from 0.02div to 4div.
8. Press the Create Mask softkey to create the mask. Whenever the Create Mask
softkey is pressed the old mask is erased and a new mask is created.
9. Press the UP softkey to return to the PASS/FAIL function menu.
10. Press the Msg Display to select On or Off to turn on or off the message display.
When On is selected, the test result will be displayed in the upper-right
message box of the screen.
F: The frame number of the channel waveform that fails to match the
mask.
P: The frame number of the channel waveform that matches the mask.
T: The frame number of the channel waveform that have been acquired. It
is the total of P and F.
11. Press the Next Page softkey to go to the second page of the PASS/FAIL function
menu.
12. Press the Stop on Fail softkey to select On or Off to turn on or off the function.
On: When failed waveforms are detected, the oscilloscope will stop the
test and enter the STOP state. At this point, the results of the test remain
the same on the screen (if display is turned on) and only one pulse is
output from the [TRIG OUT] connector (if enabled) at the rear panel.
Off: The oscilloscope will continue with the test even though failed
waveforms are detected. The test results on the screen will update
continuously and a pulse will be output from the [Trigger Out] connector at
the rear panel each time a failed waveform is detected.
13. Press the Output softkey to select turn on or off the sound.
: When the failed waveforms are detected, the beeper sounds.
: The beeper will not sound even if the failed waveforms are detected.
14. Press the Next Page softkey to return to the first page of the PASS/FAIL function
menu.
15. Press the Operate softkey to perform the test.
: Current state is stop; press the softkey to start the pass/fail test.
: Current state is running; press the softkey to stop the pass/fail test.
Save and Recall Test Mask
Users can save the current test mask to the internal flash memory or external USB
flash device. The file extension used for test mask files is “*.msk”.
Save Test Mask to Internal Memory
Loading ...
Loading ...
Loading ...