Figure 25 Slope Trigger
7. Press the Limit Range softkey; then turn the Universal Knob to select the
desired slope condition, and push down the knob to confirm.
<= (less than a time value): Trigger when the positive or negative slope
time of the input signal is lower than the specified time value.
>= (greater than a time value): Trigger when the positive or negative slope
time of the input signal is greater than the specified time value.
[--,--] (within a range of time value):Trigger when the positive or negative
slope time of the input signal is greater than the specified lower limit of
time and lower than the specified upper limit of time value.
--][-- (outside a range of time value): Trigger when the positive or negative
slope time of the input signal is greater than the specified upper limit of
time and lower than the specified lower limit of time value.
Coupling and noise reject can be set in slope trigger, see the sections "Trigger
Coupling" and "Noise Rejection" for details.