Manual - Page 191

For SDS2202X-E.

Loading ...
Loading ...
Loading ...
Internal Save and Recall
Internal save and recall support Setups in Save/Recall. In the following part, the
save and recall method and procedures are introduced.
Save the specified oscilloscope setting in internal memory.
1. Connect the signal to the oscilloscope and obtain a stable display.
2. Press Save/Recall button on the front panel to enter the SAVE/RECALL
function menu.
3. Press Save softkey to enter the SAVE function menu.
4. Press the Type softkey and then turn the Universal Knob to select Setups;
and then press the knob to confirm.
5. Press the Save To softkey to select Internal to save the current setup of the
oscilloscope to the internal memory.
6. Press the Setup softkey button; and then turn the Universal Knob to select
the location to save. The internal memory can save as many as 20 setup
files, from No.1No.20.
7. Press the Press To Save softkey to save the current setup to the appointed
location. After a few seconds, it will pop-up the message Saved to
internal file NO.1!
Load the specified type of file in internal memory.
If want to recall the setup after having finished the steps above, please do the
following steps:
1. Press Recall softkey enter the RECALLING function menu.
2. Press Type softkey and then turn the Universal Knob to select Setups; and
then press the knob to confirm.
3. Press the Recall from softkey to select Internal.
4. Press Setup softkey button; and then turn the Universal Knob to select the
location to recall. The internal memory can recall as many as 20 setup files,
from No.1No.20.
5. Press the Press To Recall softkey to recall the current setup form the
appointed location. After a few seconds, it will pop-up the message
Recalled file successfully!