Relative mV MEASUREMENTS
To measure the Relative mV of a sample:
• Press
then (select channel first for
HI5222).
• Verify if a current calibration has been made.
If required, conduct the single point Rel mV calibra-
tion. Verify the tip of the electrode is immersed into
the known solution or ORP standard.
Press
. Use and keys
set the standard value. Press to store the
calibration.
• Press
then (select channel first for
HI5222).
Place calibrated sensor tip into the sample to be
analyzed. The instrument will display the measured
Relative mV value on the LCD, together with a short
GLP information about the last calibration or Offset:
0.0 mV no Rel mV calibration was performed.
Notes: If the ORP sensor is not in solution or the
measured mV potential is out of range, “----” will be
displayed on the LCD.
mV & Relative mV MEASUREMENTS