41
Press
while in ISE Measure mode.
Press
.
Use
or to select the Calibration option.
• Press
and use or to highlight the Manual Entry option.
• Press
or to select the desired option.
Edit Custom Standards
Use Edit Custom Standards function to add additional
ISE standard values. Up to five custom standard values
can be added. Set Electrode Type and Concentration Unit
prior to adding these standards.
To edit/set the Custom Standards:
Press
while in ISE Measure mode.
Press
.
• Use
or to select the Calibration
option.
Press
and use or to
highlight the Edit Custom Standards option.
If you want to disable the custom standard, press
. A pop-up menu will be displayed
asking for confirmation. Press
to confirm (the custom standard value will turn to “----”)
or press to cancel the operation.
Use
key to select the next custom standard to be set.
Press
to return to Edit Custom Standard options.
ISE SETUP (HI5222 only)
Edit Standard Group
If a Group Standard was selected in the parameter Manual Entry, this parameter is used to create
your group of standards. If the Standard Group already contains five ISE standards, at least one
ISE standard has to be removed in order to add another standard.
To edit/set the Standard Group:
Press
while in ISE Measure mode.
Press
.
To set the Manual Entry: