To set the Alarm Limits:
Press
while in pH Measure mode.
Press
.
• Use
or to select the Alarm option.
Press
and use or to
highlight the Alarm Limits option.
Press
and use / to select
next/previous entry to be edited.
Press
and use or to set the
desired value, then press
to save the modified
value.
• Press
to return to the Alarm options. The
modified option is saved automatically.
Isopotential Point
This option allows the user to edit the isopotential point of the electrode used for pH measurements.
The isopotential point is the mV reading for an electrode at which temperature has no effect on the
measurement. The ideal electrode has an isopotential point of 0.0 mV and 7.00 pH, while an actual
electrode typically deviates slightly from the ideal values.
If the actual isopotential pH for an electrode is known, it can be set by accessing this option.
Note: If the isopotential point has been modified, recalibration must be performed.
To set the Isopotential Point:
Press
while in pH Measure mode.
Press
.
• Use
or to select the Isopotential
Point option.
• Press
and set the desired isopotential pH value
using or .
• Press
to save the current value or press
to cancel operation.
pH SETUP