• Select Load Profile to select a profile to use from the list of saved profiles. Highlight the desired
profile and press .
• Select Delete Profile to remove a selected profile from the saved list.
Highlight the profile and press
.
Temperature
The temperature has a direct influence on pH. This option allows the user to choose the temperature
source and units, as well as the desired manual temperature for manual temperature compensation
mode.
Temperature Source (HI5222 only)
If using a temperature probe, Automatic Temperature Compensation will be performed relative to the
displayed temperature, with the “ATC” indicator displayed on the LCD. For the HI5222, one temperature
probe can be used for both measurement channels if desired. Select the source by selecting Manual,
Channel 1 or Channel 2 . If no temperature probe is detected, Manual Temperature Compensation
will be performed, with the “MTC” indicator on the LCD.
Temperature Unit
The desired temperature unit can be chosen (Celsius, Fahrenheit or Kelvin degrees) and the meter will
automatically make the conversion for the selected unit.
If no temperature probe is connected, the desired temperature can be set manually. The default setting
is 25.0 °C. If the measured temperature is different, the value can be manually adjusted to obtain
an accurate pH measurement.
To set one of the Temperature options:
Press
while in pH Measure mode.
Press .
• Use or to select the Temperature
option.
• Press and use or to highlight
the desired Temperature option you wish to modify.
• Press
and use or to highlight
the desired option (for Temperature Source & Unit
options) or use or to adjust the
temperature value between the displayed limits (for
Manual Temperature option).
pH SETUP