Note: If an ISE calibration was performed and a
different Ion Selective Electrode is selected (standard
or custom), a warning message appears on the LCD
informing the user to perform a new calibration or to
select the previous ISE in order to perform accurate
measurements.
Concentration Unit
Select the desired concentration unit for the measured ion or chemical compound. The available
concentration units are: ppt, g/L, ppm, mg/L, µg/mL, ppb, µg/L, mg/mL, M, mol/L, mmol/L,
%w/v and User (custom unit).
To set the Concentration Unit:
Press
while in ISE Measure mode.
Press
.
Use
or to select the Concentration
Unit option.
Press
and use or to
highlight the desired option.
• Press
to confirm your selection or press
to cancel operation.
Sample ID - See pH Setup section.
Stability Criteria - See pH Setup section.
Log - See pH Setup section.
Note: The Logging Data Configuration option includes also the Ion Constants parameter. If you
want it to appear in the log reports, it must be enabled.
Alarm - See pH Setup section.
Note: The Alarm Limits (Low and High) are set in the selected concentration unit of the measured
ISE SETUP (HI5222 only)