27
Set Reminder Period
Schedule the calibration reminder timing with this option
(verify Daily or Periodic is set for Calibration Reminder).
If a Daily reminder is desired, set the time of day you
wish the reminder to occur.
If a Periodic reminder is desired, schedule time in days,
hours and/or minutes after the last calibration for the
reminder to occur.
To set the Reminder Period:
Press
while in pH Measure mode.
Press
.
• Use
or to select the Calibration
option.
• Press
and use or to highlight the Set Reminder Period option.
Press
and use / to select next/previous entry to be edited.
• Press
and use or to set the desired value, then press to save
the modified value.
• Press
to return to the Calibration options. If the Saving Confirmation is enabled, press
to accept the modified option, to escape without saving or to return
to the editing mode. Otherwise, the modified option is saved automatically.
Clear Calibration
This feature deletes the pH electrode calibration for the attached electrode and channel (HI5222). A
default meter calibration will replace the actual electrode calibration until a new electrode calibration
is made.
• Press
and use or to highlight the Calibration Reminder option.
• Press
and use or to highlight the desired option.
• Press
to confirm your selection or press to cancel operation.
pH SETUP