Portable Calibration Baths
Specifications
5
Input Module Option Specifications
The Product specifications describe the Absolute Instrumental Uncertainty of the
Product at 95 % level of confidence (coverage factor k = 2) within one year from
calibration. The Input Module specifications include calibration uncertainty,
linearity, repeatability, resolution, stability, and environmental temperature effects
from 13 °C to 33 °C.
Function Range
Accuracy:
Maximum Permissible Error
Reference Input Resistance
0 Ω to 42 Ω 0.0025 Ω
42 Ω to 400 Ω 0.006 %
Reference Input Temperature -25 °C to 250 °C 0.007 % + 0.015 °C
Resistance Sensing Current 1 mA 8 %
DUT 4-wire Resistance
0 Ω to 31 Ω 0.0025 Ω
31 Ω to 400 Ω 0.008 %
DUT 3-wire Resistance Accuracy 0 Ω to 400 Ω 0.12 Ω
Thermocouple mV -10 mV to 100 mV 0.025 % + 0.01 mV
Reference Junction Temperature 0 °C to 40 °C 0.35 °C
Thermocouple Temperature
-25 °C to 250 °C
J: 0.44 °C
K: 0.49 °C
T: 0.53 °C
E: 0.44 °C
N: 0.57 °C
M: 0.48 °C
L: 0.42 °C
U: 0.48 °C
0 °C to 250 °C
R: 1.92 °C
S: 1.88 °C
C: 0.84 °C
D: 1.12 °C
G/W: 3.34 °C
mA 0 mA to 22 mA 0.02 % + 0.002 mA
Notes
• Specifications stated in percent are percent of reading.
• If the environment temperature is outside the Performance Environment Range but within the Operating Environment
Range, multiply the accuracy specifications by 1.5.
• Reference Input Temperature Accuracy assumes a 4-wire, 100 Ω, α = 0.00392 PRT. The specification does not include
accuracy of the thermometer (see Table 4 of this manual).
• DUT Input Resistance Accuracy specification for 2-wire sensors is 0.05 Ω plus lead wire resistance.
• Thermocouple Input Temperature Accuracy specification includes Thermocouple Input mV and Reference Junction
Temperature, combined using the root-sum-square method. The specification does not include the accuracy of the
thermocouple.