Portable Calibration Baths
Specifications
5
Input Module Option Specifications
The Product specifications describe the Absolute Instrumental Uncertainty of the
Product at 95 % level of confidence (coverage factor k = 2) within one year from
calibration. The Input Module specifications include calibration uncertainty,
linearity, repeatability, resolution, stability, and environmental temperature effects
from 13 °C to 33 °C.
Function Range
Accuracy:
Maximum Permissible Error
Reference Input Resistance
0 to 42 0.0025
42 to 400 0.006 %
Reference Input Temperature -25 °C to 250 °C 0.007 % + 0.015 °C
Resistance Sensing Current 1 mA 8 %
DUT 4-wire Resistance
0 to 31 0.0025
31 to 400 0.008 %
DUT 3-wire Resistance Accuracy 0 to 400 0.12
Thermocouple mV -10 mV to 100 mV 0.025 % + 0.01 mV
Reference Junction Temperature 0 °C to 40 °C 0.35 °C
Thermocouple Temperature
-25 °C to 250 °C
J: 0.44 °C
K: 0.49 °C
T: 0.53 °C
E: 0.44 °C
N: 0.57 °C
M: 0.48 °C
L: 0.42 °C
U: 0.48 °C
0 °C to 250 °C
R: 1.92 °C
S: 1.88 °C
C: 0.84 °C
D: 1.12 °C
G/W: 3.34 °C
mA 0 mA to 22 mA 0.02 % + 0.002 mA
Notes
Specifications stated in percent are percent of reading.
If the environment temperature is outside the Performance Environment Range but within the Operating Environment
Range, multiply the accuracy specifications by 1.5.
Reference Input Temperature Accuracy assumes a 4-wire, 100 , α = 0.00392 PRT. The specification does not include
accuracy of the thermometer (see Table 4 of this manual).
DUT Input Resistance Accuracy specification for 2-wire sensors is 0.05 plus lead wire resistance.
Thermocouple Input Temperature Accuracy specification includes Thermocouple Input mV and Reference Junction
Temperature, combined using the root-sum-square method. The specification does not include the accuracy of the
thermocouple.