6109A/7109A
Operators Manual
72
Table 12 lists calibration points. Those marked with an asterisk are also
alignment points.
Table 12. Calibration Points for Input Module Calibration
Number Terminals Configuration Quantity
Calibration
Parameter
1 *
Reference
Reference:
Resistance
0 REF1C0
2 25 (none)
3 * 100 REF1C100
4 200 (none)
5 400 (none)
6 DUT 4-wire
DUT: RTD, 4
wires,
resistance
100 (none)
7 DUT 3-wire
DUT: RTD, 3
wires,
resistance
100 (none)
8
Thermocouple
DUT:
Thermocouple,
mV
-10 mV (none)
9 * 0 mV TCC0
10 50 mV (none)
11 * 100 mV TCC100
12 *
DUT:
Thermocouple,
E
25 °C or 35 °C TCCRJ
13
4–20 mA
DUT: mA, Loop
Power OFF,
default scaling
0 mA (none)
14 * 4 mA mAC4
15 12 mA (none)
16 20 mA (none)
17 * 22 mA mAC22