6109A/7109A
Operators Manual
50
CALCulate2:SCALe:OFFSet <numeric_value>|DEF
CALCulate2:SCALe:OFFSet?
(-P models only) Set the mA scaling offset.
<n> is the reading at 4 mA. The default is 4.
This command is conditionally password protected.
CALCulate2:SCALe:UNIT <string>|DEF
CALCulate2:SCALe:UNIT?
(-P models only) Set the mA scaling unit of measurement.
<string> is an alphanumeric string containing up to 4 characters, enclosed in
quotation marks. Allowed characters are ‘A’ to ‘Z’, ‘a’ to ‘z’, ‘0’ to ‘9’, ‘/’, ‘-‘, ‘%’,
and ‘@’. The character ‘@’ is displayed as the degree symbol ° on the display.
The default setting is “mA”.
This command is conditionally password protected.
DATA:RECord[:STATe] <boolean>
DATA:RECord[:STATe]?
Turns data recording on or off.
A properly formatted memory device must be already connected to the USB host
port.
<boolean> is 1(ON) or 0 (OFF).
DATA:RECord:FILename <filename>
DATA:RECord:FILename?
Specify the name of the file to which data is recorded.
<filename> is the name of the file that can have up to 14 characters. The
extension “.txt.” is automatically appended to the filename when the file is
created.
DISPlay:BRIGhtness <numeric_value>
DISPlay:BRIGhtness?
Set the brightness of the front panel LCD.
<numeric_value> is the fractional brightness from 0 to 1. The actual brightness is
rounded to 0.4 (low), 0.7 (medium), or 1.0 (high).