Connect the Reference Probe and Test Sensor
23
Configure the RTD Type
To configure the test RTD to accurately measure the temperature or the
resistance of the RTD:
1. Push .
2. Push (Probe).
3. Push (DUT).
4. Push to edit the DUT Type.
5. Push or to move the highlight to RTD.
6. Push to save the DUT Type.
7. Push to move the highlight to the Wires control.
8. Push to enter edit mode.
9. Push or to move the highlight to the correct number of wires.
10. Push to save.
11. Push to move the highlight to the RTD Type control.
12. Push to enter edit mode.
13. Push or to move the highlight to the correct type.
14. Push to save.
15. Push (Done) to return the first screen.
When properly connected and configured, the temperature or resistance of the
test RTD shows on the display under DUT in Monitor mode.
Connect a Thermocouple
The Input Module measures temperature from a thermocouple and automatically
applies reference junction compensation. The thermocouple connects directly to
the miniature thermocouple connector labeled TC. Do not use an external
reference junction. The positive wire connects to the smaller blade of the
connector.