Portable Calibration Baths
Operation
27
Preset Setpoints
Use presets to set the setpoint to frequently-used temperatures. To define
presets:
1. Push .
2. Push (Presets) to see the list of preset setpoints.
3. Push as needed to choose one of the numbered setpoints.
4. Push (Edit) to edit the setpoint.
5. Use the number keys to type temperature.
6. Push to save the new value.
7. Push (Done) after all presets have been defined.
To recall a preset:
1. Push .
2. Push (Presets) to see the list of preset setpoints,
3. Push as needed to choose one of the numbered setpoints.
4. Push to activate the chosen preset. The Setpoint field on the display
shows that the setpoint has changed to match the selected preset.
Set the Cutout
The cutout is a safety device that prevents overheating of the Product and bath
fluid in the event of operator error or product malfunction. For each test, set the
cutout to 5 °C to 15 °C above the maximum temperature at which the Product will
operate. Do not set the cutout higher than the maximum safe temperature of the
fluid.
1. Push .
2. Push (Operation).
3. Push to edit the Cutout Temperature.
4. Use the number keys to type in a temperature.
5. Push to save the new temperature.
Stop Heating or Cooling
Push to turn off heating and cooling and the stir motor.
Select the Temperature Unit
To change the temperature to degrees Fahrenheit or degrees Celsius:
1. Push .
2. Push (Operation).
3. Push (°C or °F).
4. Push to see temperatures in the selected unit of measurement.
Note
In some regions, the Product may not have this option.