FLUKE 6109A-P-156 Portable Calibration Bath with Process Electronics, 35°C to 250°C

User Manual for the Fluke 6109A-P-156 Portable Calibration Bath with Process Electronics, 35°C to 250°C - Page 49

For 6109A-P-156.

PDF File Manual, 83 pages, Read Online | Download pdf file

6109A-P-156 photo
Loading ...
Loading ...
Loading ...
Portable Calibration Baths
Commands
45
*ESR?
Return and clear the IEEE-488.2 Standard Event Status Register.
The response is a decimal number that ranges from 0 to 255. The definition of
the bits are shown in Table 5.
Table 5. *ESR Bits
Bit Name Definition
0 OPC Operation complete
1 RQC Request control (not used)
2 QYE Query error flag (not used)
3 DDE Device-specific error flag
4 EXE Execution error flag
5 CME Command error flag
6 URQ User request (not used)
7 PON Power on
*IDN?
Return product information. The response contains the manufacturer, model
number, serial number, and firmware version numbers.
*OPC
*OPC?
Operation complete. Since all commands are sequential, this command is not
necessary for synchronization.
*OPT?
Return the product configuration.
The response is 1 if the Product is equipped with the Input Module (7109A-P or
6109A-P only). It is 0 otherwise.
*RST
Reset operational settings.
Stop temperature control
Set the setpoint to 25 °C
Set Control Sensor to Internal
Set Ramp Enable to OFF
Set Stability Beep to ON
Set Stability Window to 0.05 °C
Stop program execution