FLUKE 6109A-P-156 Portable Calibration Bath with Process Electronics, 35°C to 250°C

User Manual for the Fluke 6109A-P-156 Portable Calibration Bath with Process Electronics, 35°C to 250°C - Page 59

For 6109A-P-156.

PDF File Manual, 83 pages, Read Online | Download pdf file

6109A-P-156 photo
Loading ...
Loading ...
Loading ...
Portable Calibration Baths
Commands
55
PROGram[:SELected]:DEFine <name>[,<numeric_value>[,...]]
PROGram[:SELected]:DEFine?
Create a new temperature program and select the new program for further
operations. This command has the same function as PROGram:EXPlicit:DEFine.
PROGram[:SELected]:NAME <name>
PROGram[:SELected]:NAME?
Select a program for further operations.
<name> is the name of an existing program enclosed in quotation marks.
PROGram[:SELected]:NUMBer <variable>,<numeric_value>|DEFault
PROGram[:SELected]:NUMBer? <variable>
Set a program variable in the selected program.
<variable> is the name of a variable (see Table 5).
<numeric_value> is the value of the variable.
PROGram[:SELected]:STATe <boolean>|RUN|PAUSe|CONTinue|STOP
PROGram[:SELected]:STATe?
Change the execution state of the selected program.
<boolean> is 0 (STOP) or 1(RUN). Which parameters are allowed depends on
the present state.
*RST stops program execution.
The query returns the present state, either STOPPED, RUNNING, or PAUSED.
PROGram:MEMory:CLEar:ALL
(-P models only) Clear memory of all test reports.
This command is password protected.
PROGram:REPort:COUNt?
(-P models only) Return the number of test reports stored in memory.
PROGram:REPort:ENABle <boolean>|DEFault
PROGram:REPort:ENABle?
(-P models only) Turn Program Report on or off.
<boolean> is 0 (OFF) or 1(ON, default).
This command is conditionally password protected.