FLUKE 6109A-P-156 Portable Calibration Bath with Process Electronics, 35°C to 250°C

User Manual for the Fluke 6109A-P-156 Portable Calibration Bath with Process Electronics, 35°C to 250°C - Page 35

For 6109A-P-156.

PDF File Manual, 83 pages, Read Online | Download pdf file

6109A-P-156 photo
Loading ...
Loading ...
Loading ...
Portable Calibration Baths
Menu Guide
31
(Monitor > Recording) Filename
Specify the filename for recording data to the memory device.
The filename can have up to 20 characters. The extension “.txt” is
automatically added to the filename when the file is created.
(Monitor) View Graph
Plot temperature over time.
The graph scales automatically to best fit the data.
For the 7109A-P and 6109A-P, the , , and softkeys
select which parameter to plot—either the control sensor, reference
probe, or DUT sensor.
(Monitor) Statistics
Show moving average and standard deviation of temperature readings.
(Monitor > Statistics) Time Window
Select the time window for the moving average and standard deviation.
(Monitor > Statistics) Reset Statistics
Clear the statistical buffers and set a new starting point for subsequent
statistics.
(Monitor) Delta Temperature/DUT Reading (-P models)
Select the parameter to be shown in the DUT field on the Monitor screen.
DUT Reading is the immediate indication from the test sensor.
Delta Temperature is the difference between the DUT reading and the
reference probe temperature. Delta T is only valid if the Reference
and DUT are both configured to read temperature.
Setpoint
Use Setpoint mode to set the temperature of the Product:
1. Push to change the setpoint.
2. Type in a new temperature using the number keys.
3. Push to save the number.
4. Push again to enable temperature control.
(Setpoint) Enable/Disable
Enable or disable temperature control.
This does the same thing as when is pushed twice after , or
.
(Setpoint) Presets
Select a preset for recall or edit.
Push or to move the focus to one of the presets in the list and
to recall that setpoint.