SDS2000X Series Users Manual - Page 78

For SDS2202X.

Loading ...
Loading ...
Loading ...
SIGLENT
58 SDS2000X User Manual
If the signal you are probing is noisy, you can set up the oscilloscope to reduce the noise
in the trigger path and on the displayed waveform. First, stabilize the displayed waveform
by removing the noise from the trigger path. Second, reduce the noise on the displayed
waveform.
1. Connect a signal to the oscilloscope and obtain a stable display.
2. Remove the noise from the trigger path by setting trigger coupling to LF Reject, HF
Reject or turning on Noise Reject.
3. Set the Acquisition option to Average to reduce noise on the displayed waveform.
Trigger Type
SDS2000X provides multiple advanced trigger functions, including various serial bus
triggers.
Edge trigger
Slope trigger
Pulse trigger
Video trigger
Window trigger
Interval trigger
DropOut trigger
Runt trigger
Pattern trigger