SDS2000X Series Users Manual - Page 138

For SDS2202X.

Loading ...
Loading ...
Loading ...
SIGLENT
118 SDS2000X User Manual
To Make Cursor Measurements
1. Press the Cursors button on the front panel to enter the CURSOR function menu.
2. Press the Mode softkey and set the option to On.
3. Press the Source softkey, and then use the Universal Knob to select the desired
source. Only analog channels (CH1~4), math waveforms and reference waveforms
(REFA, REFB,REFC,REFD) that are displayed are available for cursors.
4. To make cursor measurements:
To measure the horizontal time, use the Universal Knob to move the X1 and X2
cursors to desired place. If necessary, set the cursor type to X1-X2, move X1 and
X2 cursors together.
To measure vertical voltage or current, use the Universal Knob to move the Y1
and Y2 cursors to desired place. If necessary, set the cursor type to ―Y1-Y2‖,
move Y1 and Y2 cursors together.
To adjust the transparence of the cursors message box, press the
Display/Persist button and go to the second page, press the Transparence
(20% to 80%) softkey and then turn the Universal Knob to adjust the
transparence to the desired value.
Cursor examples:
1. Use cursors to measure pulsewidth:
Figure 69: Measure Pulse Width