SDS2000X Series Users Manual - Page 143

For SDS2202X.

Loading ...
Loading ...
Loading ...
SIGLENT
SDS2000X User Manual 123
Delay Measurements
Delay measurements measure the time different between arbitrary two channels,
including 9 kinds of delay measurements.
1. Phase: The phase difference between two channels.
2. FRR: The time between the first rising edge of source 1 and the first rising edge of
source 2 of 50 voltage level.
3. FRF: The time between the first rising edge of source 1 and the first falling edge of
source 2 of 50 voltage level.
4. FFR: The time between the first falling edge of source 1 and the first rising edge of
source 2 of 50 voltage level.
5. FFF: The time between the first falling edge of source 1 and the first falling edge of
source 2 of 50 voltage level.
6. LRR: The time between the last rising edge of source 1 and the last rising edge of
source 2 of 50 voltage level. LRR: The time between the last rising edge of source 1
and the last falling edge of source 2 of 50 voltage level.
7. LRF: The time between the last rising edge of source 1 and the last falling edge of
source 2 of 50 voltage level.
8. LFR: The time between the last falling edge of source 1 and the last rising edge of
source 2 of 50 voltage level.
9. LFF: The time between the last falling edge of source 1 and the last falling edge of
source 2 of 50 voltage level.