SDS2000X Series Users Manual - Page 168

For SDS2202X.

Loading ...
Loading ...
Loading ...
SIGLENT
148 SDS2000X User Manual
To Set and Perform Pass/Fail Test
Do the following steps to set and perform pass/fail test:
1. Press the Utility button on the front panel to enter the UTILITY function menu.
2. Press the Next Page softkey to go to the second page of the UTILITY function menu.
3. Press the Pass/Fail softkey to enter the PASS/FAIL function menu.
4. Press the Enable Test softkey to select On to enable the pass/ fail test.
5. Press the Source softkey to select the desired channel.
6. Press the Mask Setting softkey to enter the MASK function menu.
7. Press the X Mask or Y Mask softkey; and then turn the Universal Knob to select the
desired value. The range is from 0.02div to 4div.
8. Press the Create Mask button to create the mask. Whenever the Create Mask
softkey is pressed the old mask is erased and a new mask is created.
9. Press the UP softkey to return to the PASS/FAIL function menu.
10. Press the Msg Display to select On or Off to turn on or off the message display.
When On is selected, the test result will be displayed in the upper-right message box
of the screen.
Fail: Msg Display shows the number of times the oscilloscope trace has triggered the
trace gone outside of the mask area.
Pass: Msg Display shows the number of times the oscilloscope trace has triggered the
trace stayed within the mask area.
Total: the frame number of the channel waveform that have been acquired. It is
the total of Pass and Fail.
11. Press the Next Page softkey to go to the second page of the PASS/FAIL menu.
12. Press the Fail softkey to select On or Off to turn on or off the function.
On: when failed waveforms are detected, the oscilloscope will stop the test and
enter the STOP state. At this point, the results of the test remain the same on the
screen (if display is turned on) and only one pulse is output from the [TRIG
OUT]connector (if enabled) at the rear panel.
Off: the oscilloscope will continue with the test even though failed waveforms are
detected. The test results on the screen will update continuously and a pulse will
be output from the [Trigger Out] connector at the rear panel each time a failed
waveform is detected.
13. Press the Output softkey to select turn on or off the sound.
: When the failed waveforms are detected, the beeper sounds.
: The beeper will not sound even if the failed waveforms are detected,
14. Press the Next Page softkey to return to the first page of the PASS/FAIL menu.
15. Press the Operate softkey to perform the test.
: Current state is stop; press the softkey to start the pass/fail test.
: Current state is running; press the softkey to stop the pass/fail test.