SDS5000X Series Digital Oscilloscope User Manual
97 / 2 3 6 W W W. S I G L E N T. C O M
If the upper trigger level is within the waveform amplitude range while
the lower trigger level is out of the waveform amplitude range, the
oscilloscope will trigger on the rising edge only.
If the lower trigger level is within the waveform amplitude range while
the upper trigger level is out of the waveform amplitude range, the
oscilloscope will trigger on the falling edge only.
To set window trigger via the Absolute window type
Refer to "Adjust Upper/Lower Level" in the section "Slope Trigger".
To set window trigger via the Relative window type
When the window trigger type is set to
"Relative", touch the trigger descriptor
box. The pop-up menu will show two
user defined parameters: "Level +/-
Delta" and "Level Center".
The above two parameters can be set
in the following two ways:
1. Select the parameter in the Level +/-Delta area of the quick menu, then
set the parameter value by the virtual keypad, or the Level knob on the
front panel. To set the center level is the similar.
2. Directly use the Level knob on the front panel. Press the knob to switch
between Level +/-Delta" and "Center Level", and rotate it to set values.