SDS5000X Series Digital Oscilloscope User Manual
66 / 2 3 6 W W W. S I G L E N T. C O M
and may be less than memory depth. The actual sample points information can
be obtained in the timebase descriptor box (see the section "Timebase and
Trigger " for details).
The maximum memory depth in single channel mode is 2 times that of the dual
channel mode, as the following table:
Single Channel
Mode
Dual Channel
Mode
250 k
125 k
1.25 M
625 k
2.5 M
1.25 M
12.5 M
6.25 M
25 M
12.5 M
125 M
62.5 M
250 M
125 M
Single channel mode: Only one of C1/C2 is turned on, and only one of C3/C4
is turned on.
Dual channel mode: Both C1/C2 are turned on, or both C3/C4 are turned on.
Interpolation-- At small timebase settings, the number of original points on the
screen may be less than the number display pixels in the grid area, so
interpolation is necessary to display continuous waveform. For example, at 1
ns/div timebase and 5 GSa/s sample rate, the number of original points is 50,
but the grid area includes 1000 horizontal pixels. In this case, the oscilloscope
needs to interpolate the original points by 20.