27.11 Do Self Cal
The self-calibration program can quickly calibrate the oscilloscope to reach the
best working state and the most precise measurement. It is recommended to
perform a self-calibration if the change of ambient temperature is more than
5 .
Note: Make sure the oscilloscope has been warmed up or operated for more
than 30 minutes before the self-calibration.
Please do the self-calibration as follows:
1. Disconnect everything from all inputs.
2. Operate Utility>Do Self Cal, and it will pop up the following dialog.
Select Continue to start the self-cal program.
3. The oscilloscope will not response to any operate until the self-cal is
finished. After the self-cal is completed, touch the screen or press any
button to exit.