Configuration
Integration Function Configuration 7
7-21
In the Integration Setup menu, define the following settings:
Line Function
En/Start Enable integration function/ set start conditions
Clr/Stop Configure data reset / set stop conditions
Val 1..3 Select first three values
Val 4..6 Select next three values
Menu Integration Setup is displayed on the screen.
Select Integration Value
esn029.eps
1. Select with the cursor in line Val 1..3 or Val 4..6 a value and press Enter.
A dialog window showing the selectable values is displayed.
2. Move the Cursor in the window to the wanted value and press Enter to confirm.
The parameter is now shown on the display.
3. Configure the other values accordingly.
4. Save the configuration settings by pressing the SAVE function key.
Configure Status
In this menu you can enable / disable the integration function. Also the way of clearing
the values can be configured. This is done in the Integration Setup menu at column State.
Line Settings Description
En ON Integration function active
OFF Integration function inactive
Clr MAN Clear manual
AUTO Auto clear at start
Menu Integration Setup, first field column State is highlighted.
1. Press Enter, select ON and confirm with Enter.
The integrations function is now enabled. If you want to disable it, select OFF
and press Enter to confirm.
2. Save the configuration settings by pressing the SAVE function key.
Note
The integration function is enabled (ON) in the factory configuration 1:W3.