Configuration
Set Up for Measuring 7
7-3
Set Up for Measuring
Prior to measuring, you must configure the default settings, adjust channels, measuring
ranges and times, and synchronize current and voltage sources.
If you wish to reapply certain settings at a later stage, you must save the configuration.
You have the option to save up to 15 user-defined configurations, which are
automatically assigned the names 10:USER to 24:USER.
Configuration
When first switching on the Power Analyzer, factory configuration 1 and 2:W2 is used.
This configuration is suitable for measurements with fundamentals below 100 Hz
(average time 300 ms, synchronization source U1, low-pass filter 100 Hz).
Mode W3 is the standard configuration for three phase, 3-wattmeter measurement and
single phase measurement. Channels 1, 2, and 3 are used for three-phase measurements in
a 3-channel instrument, considering three-channel average values and totals. In 4-channel
instruments, channel 4 acts as an independent single channel. In 6-channel
configurations, the W3 connection operates as two independent three-phase systems.
In 4- or 6-channel instruments where the W2 configuration is selected, the channels
higher than 3 operate as the W3 configuration.
Note
You have the option to modify the settings for configuration 1:W3. If you
wish to save the new settings, you must do this in a new configuration.
Default configuration 1 and 2:W2 cannot be overwritten. You may save new
settings in the process or at the end of the configuration procedure. Settings
that have not been saved are lost when the device is switched off or when a
different configuration is loaded.
You may:
Modify configuration 1:W3 loaded upon startup of the device
Load an existing configuration
Create a new configuration
Delete or modify an existing configuration
The Power Analyzer features the configuration menus listed in Table 7-1.
Table 7-1. Configuration Menus
Configuration menu Description
General Setup Interfaces, printer output
Timing and Sync Setup Average time and synchronization
Clock Setup Date and time
Current Channel Setup Current channels 1 through 6
Voltage Channel Setup Voltage channels 1 through 6
Motor / Generator Setup PI1 process interface inputs
Analog Output Setup PI1 process interface outputs
Integration Setup Integration function / energy
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com