Technical Data
Technical Data Fluke NORMA 4000/500011
11-3
Technical Data Fluke NORMA 4000/5000
General Technical Data
NORMA 4000 NORMA 5000
Compact system With 1 to 3 phases up to 6 phases
Continuous averages
Interface Commands SCPI Vers.1999.0; legacy emulation of D5255 selectable
Housing Protection Class 1 metal housing, IP 40
Weight 5 kg (11 lb) 7 kg (15 lb)
Dimensions (W,H,D)
237.0 mm (9.3 in.), 150.0 mm (3HU)
(5.9 in.), 315.0 mm (12.4 in.)
447.0 mm (17.6 in.), 150.0 mm (3HU)
(5.9 in.), 315.0 mm (12.4 in.)
Display
145 mm (5.7 in.), 320 x 240 pixel; background illumination and contrast
adjustable
Operation Membrane keyboard, with cursor, function keys and direct functions
Mains connection 85 to 264 V AC (47 to 440 Hz) 120 to 300 V DC, Euro plug with switch
approximately 40 VA approximately 65 VA
Measuring terminals
4 mm safety sockets, 2 each / input; (for current inputsoptional binding post)
shunt connection via BNC socket
Calibration interval 2 years
Reference Conditions
Temperature 23 °C ±1 °C (71.6 °F to 75.2 °F)
Humidity < 60 % r.H.
Power supply 115 V / 230 V ±10 %
Power frequency 50 Hz / 60 Hz
Warm up period > 30 minutes
Ambient Conditions
Operating temperature range +5 to +35 °C (+41 °F to +95 °F)
Storage temperature range 20 to +50 °C (4 °F to +122 °F)
Climatic class B2 (according to IEC 60654-1)
Relative humidity maximum 85 %, noncondensing
Altitude Below 2000 m
Standards
Electrical safety
EN 61010-1/ 2. edition 1000 V CAT II (600 V CAT III)
Degree of pollution 2, Protection class I
EN 61558 for transformer
EN 61010-2-031/61010-2-032 for accessories
Electromagnetic compatibility
Emission IEC 61326-1, class B
Immunity IEC 61326-1 / industrial locations
Test voltages
Mains input housing (earth ground connector) 1.5 kV ac
Mains connection measuring inputs 5.4 kV ac
Measuring inputs - housing 3.3 kV ac
Measuring inputs – measuring inputs 5.4 kV ac