Configuration
Configure Current and Voltage Channels 7
7-15
Configure Input Range
You have the option to select automatic range configuration for the connected current
source (Auto). Alternatively, you can configure the range manually (Range).
With automatic configuration, the Power Analyzer determines and selects the correct
range for the connected current source.
Automatic Range Adjustment (Auto)
First field in column Auto is highlighted.
1. Press Enter.
2. Select ON and confirm by pressing Enter.
The selected settings are shown in column Auto.
3. If you wish to configure all three current channels in this way, press Set All.
All channels are now set to ON.
4. Save the configuration settings by pressing the SAVE function key.
Manual Range Adjustment (Range)
To manually configure the range for I1, enter the range in amperes or, if shunts are used,
in volts.
First field in column Auto is highlighted.
1. Press Enter, select OFF and confirm by pressing Enter.
Automatic range adjustment is now disabled.
2. Move the cursor to the value in column Range and press Enter.
3. Select a value in amperes; if you use a shunt, select a value in volts.
Note
When a value in volt is entered, automatic configuration (Auto) is set to Off.
Below Scale, option menu A/V is displayed.
1. Confirm by pressing Enter.
The settings are shown in column Range. Off is displayed in column Auto.
2. To configure all three current channels in this way, press Set All.
3. Save the configuration settings by pressing the SAVE function key.
Configure Scale
If you intend to use a shunt or a probe, you must adjust the scale for the output of the
measuring signal.
Note
The correct parameter settings are shown on the shunt or probe type plate.