Configuration
Load Configuration 7
7-5
Load Configuration
Note
If you have not set up and saved a new configuration before, you are
currently working with one of the predefined configurations, 1:W3 (factory
default) or 2:W2.
Load Configuration (Optional)
1. Proceed as described in Chapter 7, "Configure Data Transfer to Printer and PC.”
2. Press function key LOAD.
A list showing all existing configurations is displayed.
3. Select a configuration and confirm by pressing Enter.
The name of the loaded configuration, for example, 10:USER, is displayed in
menu item General Setup. W2 configurations are marked as “User2.”
Modify Loaded Configurations
To modify the loaded configuration, proceed as described in the following sections.
Configure Method
Mode W3 is for standard power measurement as single phase or 3-phase. Mode W2 is for
the 2-wattmeter method (Aron or Blondel) in 3-wire/3-phase power systems. For more
details, see the different applications and connection diagrams in Chapter 7.
Phase-to-phase voltage (Pc
Δ) or phase voltage (PcY) is used for transformer
measurement. The “Corrected Power” is a useful value for measurement of transformer
losses. The calculation, Active Power ÷ Corrected Power, has to be done depending on
the type of the transformer by using Pc
Y or PcΔ (see Chapter 10). In W2 mode, it is fixed
to Pc
Δ.
Configure Data Transfer to Printer and PC
If you wish to use an internal or external printer, or if you intend to connect a PC, you
must configure the parameters for the data exchange. This procedure consists of the
following steps:
Configure external printer
Configure interface to PC
Configure RS232
Configure IEEE488 device address
Configure network (LAN) addresses and protocol
Note
The actual selected interface is displayed in the Information row (see
Chapter 3, “Operating Controls and Display”):
RS RS232, GP IEEE488, EN Ethernet, US USB