NORMA 4000/5000
Operators Manual
8-22
To adjust the scale, use the cursor keys:
Cursor Key Function
Left or right Shift frequency axis
Up or down Change frequency axis zoom factor
Enter Confirm settings
Esc Abort scaling and restore previous setting
3. Adjust the scale of the scale of the axis, using the cursor keys, and press Enter or
Esc.
The graph with the adjusted axis is shown.
View Details of a Measured Value
1. Press function key Detail.
The details of measured value U1 (voltage) are displayed.
esn105.eps
2. Press function key Detail again.
The details of measured value I1 (current) are displayed.
3. Press function key Detail again.
The details of measured value P1 (power) are displayed.
4. Press function key Detail again.
The details of measured value U12 (phase-to-phase voltage) are displayed.
5. To return to the overview of measured values of the selected channel, press
function key Detail again (or Esc from any of the previous detail screens).
Set Frequency Range
The default frequency range is set to a maximum of half the sampling frequency.
Note
For signals with a lower frequency (for example, 10 Hz), the frequency
range must be adjusted; otherwise, the measurements would be inaccurate.