NORMA 4000/5000
Operators Manual
5-6
Measurement with Shunt
The connecting leads to the shunts should be as short as possible in order to prevent
interference and noise voltages.
XW Warning
To avoid possible electric shock or personal injury:
Do not touch sensing terminals. The sense terminals at the
shunts are powered with the same voltage as the power
connections.
Shunts are not isolated. Never touch the sense terminals at
the shunts.
Risk of injury when touching connections, internal circuits,
and measuring devices that are not earthed.
Always adhere to the instructions regarding the connection
sequence (see Chapter 5, “Connecting Sequence”).
Where the current to be measured exceeds the rating of the direct current connection, an
external triaxial shunt should be used, see Figure 5-3. Fluke triaxial shunts are
recommended as they provide high accuracy across the full frequency range. Standard
linear shunts may produce excessive errors due to the possible presence of high
frequency that components experience with electronic loads. The NORMA internal shunt
is optimized for such loads.
HI
VOLTAGE
1000 V max
LO
PROBE
EXT.SHUNT
10 V max
HI
CURRENT
10 A max
LO
ALL INPUTS MAX 1000V CATII TO
PROBE
EXT.SHUNT
10 V max
HI
CURRENT
10 A max
LO
HI
VOLTAGE
1000 V max
LO
123
Max. 10 m
Load
HILO
L1
N
Guard
*1
V
esn010.eps
Figure 5-3. Measurement with Shunt