Replacement of Battery Pack
XW Warning
Disconnect all the sensors from the instrument’s input
connectors.
Disconnect the instrument from the power supply.
Do not short circuit the terminals of the battery pack.
For replacement of the battery pack, use the original
spare parts only (PN 2540406).
Note
Always adhere to the applicable statutory regulations for recycling and
waste disposal.
Procedure:
1. Locate the battery compartment on the backside of the instrument.
2. Remove the screw of the lid with a screwdriver (Pozi-drive).
3. Unlock and remove the connector cable.
4. Replace the battery pack by an original spare part (PN 2540406) using the
attached strip.
5. Connect the cable to the plug of the instrument.
Note
Note the polarity of the plug and the locking mechanism.
Decommissioning and Disposal
Shutting Down
1. Ensure that all the devices connected to measuring circuits are disconnected from
the measuring circuits.
2. Switch off the Power Quality Recorder.
3. Disconnect the plug from the mains socket.
4. Remove all the connected devices.
5. Secure the unit against inadvertent switching on.
6. Ensure that the operating manual is kept near the device.