Maintenance
Introduction 4
4-3
Introduction
The instrument itself is maintenance-free.
Maintenance of Battery Package
Note
We recommend carrying out a forced battery discharge at regular time
intervals (no longer than 3 months) to maintain the battery capacity as long
as possible. It is recommended to replace the battery every 2 years to
ensure full operational readiness for UPS.
Procedure:
1. Connect the instrument to mains.
2. Set the mains switch to the I-position.
3. Wait until the POWER LED is on.
4. Disconnect the power supply.
5. Wait until the POWER LED goes off.
6. Set the mains switch to the 0-position.
7. Wait until the LEDs POWER and UPS are flashing rapidly.
8. Within 3 second set the mains switch to I-position again.
The battery package will be discharged completely when:
LED POWER is OFF
LED UPS is flashing slowly
LEDs MEMORY LEVEL show flashing light, the number of LEDs lighting up
indicates the remaining time period for discharging in minutes (e.g. 5 LEDs
means that the discharging will last for appr. 5 minutes)
Afterwards the instrument is turned off automatically
Note
For terminating the forced discharge mode at any time, connect instrument
to mains or set the mains switch to the 0-position.
Cleaning
The device can be cleaned with an Isopropanol impregnated cloth.
W Caution
Do not use abrasives or other solvents.