SDG1000 User Manual 67
Figure 2- 59 Table 2- 30 Menu Explanations of Output Setting (Page 1/2)
1. To Set the Output Load
For the [Output] connector on the front panel, the generator has a built-in
50Ω series impendence. If the actual load does not match the set one,
the displayed amplitude and offset will be incorrect. This function is used
to match the displayed voltage with the expected one.
Steps for setting the load:
Press Utility Output Setup Load, to enter the following interface.
Please note that the load parameter shown on the right
bottom is the default setting when the power is on or the pre-set load
value. If the current value is valid for the output, then current value will be
used.
Function
Menu
Settings
Explanation
Load
HighZ
Set the load connected to the Output
Connector;
Set the load connected to the Output
Connector to be HighZ.
Normal
Invert
Normal output;
Inverse output.
Coupling
ChCopy
Copy one channel to the other
Done
Finish operation.