Siglent SDG1032X 30MHz 2 Channel Function/Arbitrary Waveform Generator

Manual - Page 38

For SDG1032X.

PDF File Manual, 133 pages, Read Online | Download pdf file

SDG1032X photo
Loading ...
Loading ...
Loading ...
SDG1000 User Manual 27
Figure 2- 10 Table2- 3 Menu Explanations of Ramp Waveform
To Set the Symmetry
1. Press Ramp Symmetry, to set the symmetry.
The symmetry shown on the screen when the instrument is powered is
the default value or the set value beforehand. When changing the
function, if the current value is valid for the new waveform, it will be used
sequentially.
2. Input the desired Symmetry
Function
Menu
Settings
Explanation
Freq/
Period
Set the signal frequency or period;
The current parameter will be switched at
a second press.
Ampl/
HLevel
Set the signal amplitude or high level;
The current parameter will be switched at
a second press.
Offset/
LLevel
Set the signal offset or low level;
The current parameter will be switched at
a second press.
Phase/
EqPhase
Set the phase of the signal;
The current parameter will be switched at
a second press.
Symmetry
Set the symmetry for ramp waveform.
Term Explanation:
Symmetry: The percentage that the rising period takes up the whole Period.
Input Range: 0~100%.
Loading ...
Loading ...
Loading ...